◆Mirus™ 1.0 Nanopump Performance Specifications:
Shear Stress Range*: 0.05 – 20 dyne/cm2
Volumetric Flow Rates: 100nL/minute - 20 μL/minute (100 μL syringe)
Sample Volume Increments: Freely adjustable
Linear Velocity Range**: 10 μm/s to 10cm/s
Flow Direction: Reversible
Sample Volume Aspiration Accuracy: ±1%
Shear Stress Accuracy: ±0.5%
Sample Volume Aspiration Precision: <1% CV
Shear Stress Precision: <0.5% CV
*1mL syringe for high shear stress > 10 dyne/cm2.
**Given for the flow of distilled water in a Vena8TM biochip microcapillary with dimensions: 400 μm (W) x 100
μm (D) x 20mm (L).
◆Mirus™ 1.0 Nanopump Technical Specifications:
Software control: Integrated FlowAssay software facilitates pumping of
cell suspensions through biochips.
Hardware control: PC requirements: Pentium II microprocessor or higher, 128MB of RAM,
400MB Hard disk, RS232 port (COM port), USB port,
CD-ROM drive for FlowAssay software installation,
Windows 2000 ME, XP Operating systems.
Dimensions: 220mm (W) x 175mm (D) x 220mm (H)
Weight: Approx. 5kg
Power Requirements: 110/220V 50/60Hz